Material & Equipment Management

Chapter 8
Equipment Installation, Utilization and Life-Cycle Management

Post Graduate Program in Health Care Management
Material & Equipment Management   •   Post Graduate

Equipment Installation, Utilization and Life-Cycle Management

Introduction and Context

Medical equipment management extends beyond purchasing. A device creates clinical value only when it is correctly installed, safely commissioned, competently used, maintained, calibrated, documented, upgraded when appropriate, and retired responsibly. In health-care organizations, equipment failure can interrupt services, increase clinical risk, create regulatory exposure, and raise total cost of ownership.

Life-cycle management is therefore a coordinated managerial and technical process covering the period from identification of need to final disposal. It integrates clinical requirements, engineering controls, finance, procurement, information systems, infection prevention, occupational safety, and environmental stewardship.

The central managerial question is not simply, “How old is the equipment?” It is: “Can this asset continue to provide safe, reliable, clinically appropriate and economically acceptable service?” Age is relevant, but risk, performance, utilization, maintainability, technological obsolescence and clinical need are equally important.

Learning Objectives

By the end of this chapter, the learner should be able to:

  1. Explain the stages of the medical-equipment life cycle.
  2. Develop site-readiness and installation plans.
  3. Distinguish acceptance testing, commissioning and routine quality assurance.
  4. Design training, asset-registration and utilization-monitoring systems.
  5. Apply principles of preventive maintenance, calibration and upgrade management.
  6. Establish evidence-based replacement priorities.
  7. Manage decommissioning, disposal, data protection and environmental responsibilities.
  8. Evaluate equipment-management decisions using clinical, technical, financial and ethical criteria.

1. The Equipment Life Cycle as a Management System

A useful life-cycle model contains seven connected stages:

  1. Needs assessment and specification – defining the clinical problem, workload, users, performance requirements and budget.
  2. Procurement and delivery – selecting a supplier, contracting for service support and controlling logistics.
  3. Site preparation and installation – ensuring that physical, utility, information-technology and safety conditions are adequate.
  4. Acceptance and commissioning – verifying that the delivered equipment conforms to contract and is safe for clinical use.
  5. Operation and maintenance – managing use, training, preventive maintenance, corrective maintenance, calibration and performance monitoring.
  6. Upgrade, renewal or replacement – deciding whether to modify, refurbish, replace or continue operation.
  7. Decommissioning and disposal – removing the equipment safely, protecting data and managing environmental and legal obligations.

These stages are interdependent. A poor specification can produce an unsuitable purchase; inadequate site readiness can delay installation; weak training can produce misuse; and incomplete asset records can cause missed maintenance and poor replacement decisions.

Life-cycle governance should assign responsibility clearly. A typical governance group includes the clinical department, biomedical or clinical engineering, facilities, information technology, infection prevention, procurement, finance, risk management and, where relevant, radiation or laboratory safety specialists.

2. Needs Assessment and Pre-Installation Planning

Before delivery, the organization should create an equipment implementation plan. The plan should identify:

  • the clinical service and intended users;
  • patient volume, acuity and expected workload;
  • the required performance and safety characteristics;
  • compatibility with existing devices and information systems;
  • room dimensions, access routes and structural constraints;
  • electrical supply, grounding, emergency power and voltage stability;
  • medical gases, water, drainage, ventilation and temperature control;
  • networking, cybersecurity and data-integration needs;
  • infection-prevention and cleaning requirements;
  • radiation shielding or other specialized protections;
  • training, maintenance, spare parts and warranty arrangements;
  • acceptance criteria, documentation and approval authorities.

A site-readiness review should be completed before shipment. “Room available” does not necessarily mean “site ready.” A CT scanner, for example, may require structural assessment, shielding, cooling, electrical capacity, controlled access, image-network integration and regulatory inspection. A dialysis system may require water-treatment validation, drainage, electrical protection and infection-control procedures.

Site-readiness responsibility matrix

DomainTypical responsible functionEvidence of readiness
Building and accessFacilities/project managementApproved room and delivery route plan
Electrical and groundingFacilities/clinical engineeringTest certificates and load verification
Medical gases or waterFacilities/technical servicesPressure, quality and safety records
Network and cybersecurityInformation technologyApproved connectivity and access controls
Infection preventionInfection-prevention teamCleaning and reprocessing procedure
Radiation or laboratory safetySafety officer/regulatorRequired approvals and shielding evidence
Clinical workflowDepartment manager/clinical leadWorkflow and staffing plan
Training and maintenanceSupplier/biomedical engineeringTraining schedule and service agreement

A formal “go/no-go” review prevents expensive equipment from being delivered to an incomplete site. Any unresolved high-risk issue should have an owner, deadline and documented mitigation.

3. Installation, Acceptance Testing and Commissioning

Installation

Installation should be performed or supervised by qualified personnel according to the manufacturer’s instructions and applicable standards. The installation record should include serial number, configuration, software version, accessories, location, utility readings, installation date and names of responsible personnel.

Installation is not complete merely because the equipment powers on. It includes physical assembly, utility connection, software configuration, network integration, safety labeling, environmental controls and verification that accessories are correctly supplied.

Acceptance testing

Acceptance testing determines whether the supplied equipment meets the purchase specification and is safe to enter service. It normally includes:

  • identity and completeness checks;
  • comparison with the purchase order and technical specification;
  • inspection for transport damage;
  • electrical and mechanical safety tests;
  • functional testing across relevant operating modes;
  • performance testing against stated tolerances;
  • verification of alarms, interlocks and emergency stops;
  • software, connectivity and cybersecurity checks;
  • availability of manuals, certificates and service information;
  • confirmation of warranty, service response and training commitments.

Acceptance criteria must be defined before testing. If criteria are created after a failure occurs, the process becomes vulnerable to bias and dispute. A failed test should lead to documented corrective action, retesting and formal authorization before clinical release.

Commissioning

Commissioning is broader than acceptance testing. It confirms that the equipment, environment, users, procedures and support systems function together in the intended clinical setting. Commissioning may include:

  • workflow simulation;
  • clinical-user competency assessment;
  • emergency procedures and downtime drills;
  • integration with electronic health records or laboratory systems;
  • infection-control validation;
  • quality-control baseline measurements;
  • review of preventive-maintenance intervals;
  • authorization by the clinical and technical owners.

Acceptance answers, “Did we receive what we purchased?” Commissioning answers, “Can this system be used safely and effectively in our service?”

A commissioning dossier should contain the signed acceptance report, test results, certificates, drawings, configuration information, training records, risk assessment, maintenance plan and release decision.

4. Training, Human Factors and Safe Use

Training is a control measure, not a ceremonial handover. It should be role-specific and address normal operation, contraindications, alarms, cleaning, emergency shutdown, troubleshooting, documentation and escalation.

A robust training program includes:

  1. identification of user groups and competency requirements;
  2. theoretical instruction on indications, limitations and risks;
  3. supervised practical training;
  4. observed competency assessment;
  5. authorization of only competent users;
  6. refresher training after defined intervals or major upgrades;
  7. retraining after incidents, repeated errors or prolonged non-use.

Human-factors design is particularly important because equipment-related incidents often arise from confusing displays, poorly located controls, alarm fatigue, incompatible connectors, unclear labeling or workflow mismatch. Users should participate in evaluation, commissioning and post-installation review.

Training records should identify the learner, equipment, content, trainer, date, assessment result and expiry or renewal requirement. “Training completed” is weaker evidence than demonstrated competence.

5. Asset Registration and Information Governance

Every significant asset should have a unique identifier linked to a controlled asset register or computerized maintenance-management system. Minimum data commonly include:

Data categoryExamples
IdentificationAsset number, manufacturer, model, serial number
Location and ownershipSite, department, room, responsible manager
Technical profileCapacity, configuration, software version, accessories
Financial informationPurchase date, acquisition cost, funding source
Risk and statusRisk class, operational status, criticality
Service historyMaintenance, breakdowns, calibration, parts and downtime
Contractual dataWarranty, supplier, service agreement and expiry
Disposal dataRetirement date, method, certificate and data-erasure evidence

Asset records must be accurate enough to support maintenance, audit, budgeting and incident investigation. Duplicate records, missing serial numbers and unrecorded transfers undermine control.

Connected equipment introduces information-governance risks. Devices may store patient identifiers, images, measurements, user credentials or network information. Before disposal, transfer or repair, the organization should apply an approved data-retention and secure-erasure process. Cybersecurity controls should include least-privilege access, patch management, vulnerability assessment, network segmentation where appropriate, backup and incident response.

6. Utilization Monitoring and Performance Management

Utilization should be assessed in relation to clinical need, capacity and service quality. A low number of examinations may indicate low demand, poor scheduling, inadequate staffing, excessive downtime or inappropriate location. A high workload may indicate efficient use—or unsafe overloading and insufficient capacity.

A basic utilization ratio is:

U = Tₐ ÷ Tₐᵥ × 100%

where:

  • U is utilization percentage;
  • Tₐ is actual productive operating time during the measurement period;
  • Tₐᵥ is available scheduled operating time during that period.

The formula assumes that available time has been defined consistently and that productive time is measured reliably. It should not be interpreted as a universal target. Planned maintenance, cleaning, setup, emergency access and staffing constraints may make some non-productive time necessary.

Other useful measures include:

  • Availability: the proportion of required time that equipment is operational and ready for use.
  • Downtime: time unavailable because of breakdown, maintenance, supply interruption or other causes.
  • Throughput: completed procedures or cases per unit of time.
  • Turnaround time: elapsed time between defined start and completion points.
  • Repeat rate: proportion of procedures repeated because of technical or operational problems.
  • Cancellation rate: proportion of scheduled procedures not completed.
  • Mean time between failures (MTBF): operating time divided by the number of relevant failures, when failure events are clearly defined.
  • Mean time to repair (MTTR): repair time divided by the number of repairs, with the organization defining whether waiting and logistics time are included.

Metrics should be interpreted together. High utilization with declining availability and increasing repeat procedures may signal overuse or inadequate maintenance. Benchmarking should account for equipment type, case mix, staffing model and service hours.

7. Maintenance, Calibration and Quality Assurance

Maintenance includes preventive, predictive, corrective and sometimes condition-based activities.

  • Preventive maintenance is scheduled to reduce the probability of failure or performance degradation.
  • Corrective maintenance restores equipment after a fault.
  • Predictive or condition-based maintenance uses measured condition, trends or diagnostics to determine intervention timing.
  • Calibration compares a measurement instrument with a traceable reference and, where necessary, adjusts or documents its error.

Calibration is not identical to maintenance. A device may be mechanically functional but measuring inaccurately; conversely, a calibrated instrument may still have a mechanical safety defect.

A calibration record should state the instrument identity, reference standard, method, environmental conditions when relevant, results, uncertainty or tolerance information where required, adjustment performed, next due date and authorized person. Calibration intervals should be risk-based and informed by manufacturer guidance, regulatory requirements, stability history, frequency of use and consequences of inaccurate measurement.

Quality assurance should combine scheduled checks with trend analysis. Repeated minor deviations may be more informative than one obvious failure. Equipment should be removed from service when it presents unacceptable safety risk, fails a critical performance test, lacks required calibration or cannot be used within approved limits.

8. Upgrade, Obsolescence and Total Cost of Ownership

An upgrade may be preferable to replacement when the core platform remains safe and supportable, the upgrade solves a defined clinical or technical need, and compatibility risks are controlled. Replacement is more appropriate when the platform is obsolete, unsupported, unsafe, uneconomical, incompatible with current systems or unable to meet clinical requirements.

Total cost of ownership includes more than purchase price:

TCO = Cₚ + Cᵢ + Cₘ + Cₜ + Cₑ + Cd − Vᵣ

where Cₚ is purchase cost, Cᵢ installation and integration cost, Cₘ maintenance and calibration cost, Cₜ training cost, Cₑ energy and consumables cost, Cd decommissioning and disposal cost, and Vᵣ residual value or recoveries. All amounts should be expressed on a comparable basis and over a defined time horizon. If costs occur in different years, discounting may be required for investment appraisal.

A low acquisition price can therefore produce a high life-cycle cost if the equipment consumes expensive supplies, has poor local service support or causes substantial downtime.

9. Replacement Prioritization

Replacement programs should be transparent, documented and clinically defensible. A useful approach combines four dimensions:

  1. Age and supportability – age, spare-parts availability, software support and manufacturer service status.
  2. Risk – potential severity and likelihood of harm if the equipment fails or performs inaccurately.
  3. Performance – availability, failure frequency, repair cost, calibration results and user-reported problems.
  4. Clinical need – service criticality, patient volume, equity, strategic priorities and availability of alternatives.

A weighted scoring model can support, but should not replace, professional judgment:

P = wₐA + wᵣR + wₚF + w𝚌C

where P is replacement priority, A, R, F and C are normalized scores for age/supportability, risk, performance and clinical need, and wₐ, wᵣ, wₚ and w𝚌 are non-negative weights whose sum equals 1. Scores must use a documented scale, such as 0 to 5. The model assumes that the dimensions are sufficiently defined for comparison and that double-counting is controlled.

For example, suppose an organization assigns weights of 0.20, 0.35, 0.25 and 0.20 to age, risk, performance and clinical need. An asset scored 4, 5, 3 and 4 respectively receives:

P = (0.20 × 4) + (0.35 × 5) + (0.25 × 3) + (0.20 × 4) = 4.10

The score is a prioritization aid, not an automatic authorization. A lower-scoring asset may still require urgent action if a regulator declares it unsafe or if a critical failure has occurred.

10. Decommissioning and Disposal

Decommissioning should be planned before the final day of use. The process includes:

  • clinical approval and service-continuity planning;
  • identification and isolation of the asset;
  • removal of patient data, passwords and network credentials;
  • decontamination and infection-control clearance;
  • disconnection of utilities by qualified personnel;
  • removal of hazardous materials, radioactive sources or pressurized components;
  • update of the asset register and financial records;
  • evaluation for redeployment, donation, resale, recycling or destruction;
  • documented chain of custody and disposal certificate.

Donation is not automatically ethical. Equipment that is obsolete, unsupported, unsafe or incompatible with the recipient’s infrastructure may transfer risk rather than create benefit. Responsible donation requires assessment of clinical suitability, training, maintenance support, consumables, regulatory requirements and total cost for the recipient.

Electronic waste, batteries, mercury-containing components, refrigerants and other hazardous materials require controlled disposal under applicable law. Environmental considerations should include energy consumption, repairability, packaging, refurbishment and supplier take-back arrangements.

Illustrative Case: Replacement of Imaging Equipment in a Regional Hospital

Background

A regional hospital operates an eight-year-old CT scanner. It serves emergency, inpatient and outpatient departments. The scanner remains clinically useful but has increasing downtime, rising repair costs and limited software support.

Problem

The hospital must decide whether to continue maintenance, purchase an upgrade or replace the scanner. The clinical department is concerned about cancelled examinations, while finance is concerned about capital cost.

Analysis

The hospital forms a multidisciplinary review team. It validates the asset register, examines 24 months of service records, interviews users and reviews patient-safety incidents. Availability has declined, repeat scans have increased, and the supplier has announced that a major component will soon be unsupported. However, demand projections show that the service will continue to require CT capacity.

The team evaluates four options: continue existing support, purchase a partial upgrade, replace with a new scanner, or outsource selected examinations. The analysis includes purchase and installation, building modifications, service contracts, radiation-safety requirements, training, downtime during transition, energy consumption and disposal.

A replacement-priority score is calculated using the organization’s documented weights. Risk and supportability receive greater weight than age because failure could delay emergency diagnosis. The team also conducts a failure-mode review and confirms that a temporary service arrangement is needed during installation.

Approach and solution

The hospital approves replacement, but only after confirming funding, site readiness and continuity arrangements. The project includes a room survey, electrical verification, network and cybersecurity review, acceptance testing, radiation-safety inspection, clinical workflow simulation and competency assessment. The old scanner is decommissioned after data erasure, decontamination and safe removal.

Outcome

The new scanner enters service with a baseline quality-control dataset and a defined dashboard covering availability, downtime, repeat examinations, turnaround time, utilization and maintenance cost. The hospital also revises its replacement plan so that major assets are reviewed annually rather than only after failure.

Learning points

The case demonstrates that replacement is a portfolio decision, not an age-triggered event. It also shows why transition planning, acceptance testing, training, asset registration and post-implementation measurement are necessary to realize the expected benefit.

Challenges and Limitations

Life-cycle management can be weakened by incomplete data, fragmented accountability, supplier dependence, limited biomedical-engineering capacity and pressure to maximize short-term utilization. Scoring models may create false precision, especially when risk and clinical need are difficult to quantify. Utilization measures can also encourage inappropriate volume if productivity is rewarded without regard to quality or patient outcomes.

Organizations should therefore combine quantitative indicators with expert review, user feedback, incident learning and regulatory requirements. Decisions should be auditable, but not mechanically determined by a spreadsheet.

Governance, Ethics and Regulation

Equipment management must protect patient safety, privacy, equity and professional accountability. Governance questions include:

  • Who is authorized to release equipment for clinical use?
  • Who can suspend use when a safety concern arises?
  • Are maintenance and calibration records complete and retrievable?
  • Are procurement decisions free from undisclosed conflicts of interest?
  • Does technology selection widen or reduce inequity between patient groups or locations?
  • Are donated or refurbished devices safe and supportable?
  • Is patient data protected during repair, transfer and disposal?

Regulatory requirements vary by jurisdiction and equipment class. Managers should consult applicable medical-device, radiation, laboratory, electrical-safety, occupational-health, waste-management and data-protection rules. Manufacturer instructions are important but do not replace the organization’s legal and professional responsibilities.

Contemporary Developments

Digital asset registers increasingly connect maintenance records, procurement data, clinical utilization and financial planning. Internet-connected devices can support remote diagnostics and predictive maintenance, but they also expand the cybersecurity attack surface. Artificial intelligence may assist failure prediction, demand forecasting and anomaly detection; however, models require validated data, monitoring for drift, transparent escalation rules and human accountability.

Sustainable procurement is also becoming more important. Hospitals are evaluating energy efficiency, modular design, repairability, reusable components, supplier take-back programs and embodied carbon alongside price and performance. The mature objective is not maximum equipment ownership but reliable clinical capability with acceptable risk and resource use.

Best-Practice Checklist

Before clinical release, confirm that:

  • the site and utilities are approved;
  • installation and acceptance tests are documented;
  • commissioning includes workflow and safety validation;
  • users have demonstrated competence;
  • the asset is registered with a unique identifier;
  • maintenance, calibration and quality-control schedules are active;
  • cybersecurity and data-protection controls are implemented;
  • downtime and escalation procedures are known;
  • performance indicators have owners and review dates;
  • replacement and decommissioning criteria are documented.

Chapter Summary

Equipment life-cycle management is an integrated system beginning with needs assessment and ending with safe, responsible disposal. Site readiness must address facilities, utilities, information technology, infection prevention, safety and workflow. Acceptance testing verifies that the delivered asset meets specified requirements, while commissioning verifies that the complete system can operate safely in its clinical context.

Training, competency assessment and human-factors design reduce misuse. Accurate asset registration supports maintenance, audit, budgeting and replacement planning. Utilization, availability, downtime, throughput, repeat rates, MTBF and MTTR provide complementary evidence, but none should be interpreted in isolation. Maintenance and calibration are distinct but related controls, and both require documented schedules and results.

Replacement priorities should combine age and supportability, risk, performance and clinical need. Weighted scoring can improve transparency, but professional and regulatory judgment remain essential. Total cost of ownership includes acquisition, installation, maintenance, training, operating, disposal and residual-value considerations. Decommissioning must address service continuity, data security, decontamination, hazardous materials, environmental responsibilities and ethical donation. Effective management ultimately links equipment decisions to safe, equitable and sustainable patient care.

Key Terms

Life-cycle management: Coordinated management of an asset from need identification through disposal.

Site readiness: Verified suitability of the physical, utility, information and workflow environment.

Acceptance testing: Formal testing to determine whether delivered equipment meets specified requirements.

Commissioning: Verification that equipment, environment, users and procedures function together safely.

Asset register: Controlled record containing identification, location, ownership and service information.

Utilization: Use of available equipment capacity during a defined period.

Availability: Proportion of required time that equipment is operational and ready.

Downtime: Period during which equipment cannot provide its intended service.

Preventive maintenance: Planned activity intended to reduce failure probability or degradation.

Corrective maintenance: Activity that restores equipment after a fault.

Calibration: Comparison of measurement performance with a traceable reference.

Quality assurance: Systematic activities that maintain and verify required performance.

Total cost of ownership: Full cost of acquiring, operating, supporting and disposing of an asset.

Obsolescence: Loss of practical suitability because of technological, support, safety or compatibility changes.

Replacement priority: Relative urgency with which an asset should be renewed or otherwise addressed.

Decommissioning: Controlled withdrawal of equipment from service.

Clinical criticality: Importance of equipment to essential patient-care functions.

Human factors: Design and organizational considerations affecting how people interact with equipment.

Predictive maintenance: Maintenance planned using condition data or failure-prediction evidence.

Cybersecurity: Protection of connected equipment, systems and data from unauthorized access or disruption.

Review Questions

Short Answer Questions

  1. Distinguish acceptance testing from commissioning.
  2. List five domains that should be included in a site-readiness review.
  3. Why is calibration not the same as maintenance?
  4. Identify four dimensions commonly used in replacement prioritization.
  5. What information should be recorded in an asset register?

Long Answer Questions

  1. Explain the complete life cycle of medical equipment and show how weaknesses at one stage affect later stages.
  2. Discuss the design of a training and competency-assessment program for newly installed equipment.
  3. Explain how utilization, availability, downtime and repeat rates should be interpreted together.
  4. Analyze the advantages and limitations of using a weighted replacement-priority score.
  5. Describe the governance, environmental and data-protection requirements of equipment decommissioning.

Analytical / Application Questions

  1. A device has high utilization but declining availability and frequent repeat procedures. Develop an investigation and intervention plan.
  2. Using the formula P = wₐA + wᵣR + wₚF + w𝚌C, compare two assets and explain why the higher-scoring asset should not automatically be replaced without professional review.
  3. A hospital is considering a low-cost donation of unsupported equipment to a rural clinic. Evaluate the proposal using safety, infrastructure, training, maintenance, equity and environmental criteria.

Multiple Choice Questions with Answers

Question 1

Which activity primarily answers the question, “Did we receive what we purchased?”

A. Acceptance testing

B. Routine utilization monitoring

C. Predictive maintenance

D. Decommissioning

Correct Answer: A. Acceptance testing

Question 2

Which item is most directly evidence of site readiness for networked medical equipment?

A. A completed staff satisfaction survey

B. Approved connectivity and access-control verification

C. A disposal certificate

D. A depreciation schedule

Correct Answer: B. Approved connectivity and access-control verification

Question 3

A device has actual productive operating time of 360 hours and available scheduled time of 450 hours. Using U = Tₐ ÷ Tₐᵥ × 100%, what is its utilization?

A. 60%

B. 70%

C. 80%

D. 90%

Correct Answer: C. 80%

Question 4

Which statement best distinguishes calibration from maintenance?

A. Calibration addresses measurement accuracy against a reference; maintenance addresses condition or function.

B. Calibration is only needed after equipment disposal.

C. Maintenance applies only to software, while calibration applies only to buildings.

D. They are identical terms for the same activity.

Correct Answer: A. Calibration addresses measurement accuracy against a reference; maintenance addresses condition or function.

Question 5

Which factor is least appropriate as a sole basis for equipment replacement?

A. Age alone

B. Clinical criticality

C. Failure and performance history

D. Supportability and safety risk

Correct Answer: A. Age alone

Question 6

In the total-cost-of-ownership expression, what does residual value represent?

A. The cost of staff training

B. The value recovered at the end of the asset’s life

C. The cost of preventive maintenance

D. The cost of room preparation

Correct Answer: B. The value recovered at the end of the asset’s life

Question 7

Which action should occur before releasing newly installed equipment for routine clinical use?

A. Deleting all maintenance records

B. Completing acceptance and commissioning activities

C. Waiting for the first failure

D. Removing the asset from the register

Correct Answer: B. Completing acceptance and commissioning activities

Question 8

Why can high equipment utilization be misleading when considered alone?

A. It always proves that equipment is obsolete.

B. It may coexist with unsafe overloading, poor availability or inadequate quality.

C. It measures only purchase price.

D. It automatically includes patient outcomes.

Correct Answer: B. It may coexist with unsafe overloading, poor availability or inadequate quality.

Question 9

Which is an appropriate decommissioning control for connected equipment?

A. Transfer the device without changing credentials

B. Remove patient data and network credentials using an approved process

C. Leave the asset active until disposal is complete

D. Ignore hazardous components if the device is old

Correct Answer: B. Remove patient data and network credentials using an approved process

Question 10

In a weighted replacement-priority model, the weights should normally be:

A. Negative and unrelated to the scoring scale

B. Non-negative and sum to 1

C. Equal to the purchase price

D. Based only on equipment age

Correct Answer: B. Non-negative and sum to 1

Material & Equipment Management
Chapter 8 of 14