Electrical Measurements & Instrumentation

Chapter 3
Analog Electromechanical Instruments

Bachelors Program in Electrical Engineering
Electrical Measurements & Instrumentation   •   Bachelor

Analog Electromechanical Instruments

Introduction

Analog electromechanical instruments convert an electrical quantity into a mechanical displacement, usually the angular movement of a pointer over a calibrated scale. Although digital instruments dominate modern measurement systems, analog mechanisms remain important in switchboards, energy meters, rugged industrial indicators, reference standards, and the historical foundations of instrumentation engineering.

The principal instrument families differ mainly in how they produce deflecting torque. Permanent-magnet moving-coil (PMMC) instruments use the force on a current-carrying conductor in a magnetic field. Moving-iron instruments use changes in magnetic reluctance. Electrodynamometer instruments use interacting magnetic fields. Induction instruments use alternating fluxes and induced currents. Electrostatic instruments use electric-field energy, while thermal instruments infer current or voltage from heating.

Accurate use requires more than knowing the operating principle. An engineer must evaluate controlling and damping torques, dynamic response, range-extension components, waveform sensitivity, loading, calibration, environmental influences, and safety.

Learning Objectives

After studying this chapter, a student should be able to:

  • explain the operating principles of major analog electromechanical instruments;
  • derive their deflecting-torque equations under stated assumptions;
  • distinguish deflecting, controlling, and damping torques;
  • calculate shunts, series multipliers, and instrument-transformer ratios;
  • assess scale shape, dynamic response, errors, and loading;
  • select an appropriate instrument for DC, AC, power, energy, high-voltage, and high-frequency measurements; and
  • apply calibration, safety, and professional measurement practices.

1. Functional Structure of an Analog Instrument

A complete indicating instrument contains four functional elements:

  1. Deflecting system: develops torque from the measurand.
  2. Moving system: includes the pointer, spindle, coil, iron vane, or other moving element.
  3. Controlling system: produces a restoring torque and establishes a unique steady reading.
  4. Damping system: dissipates mechanical energy so that the pointer settles quickly without excessive oscillation.
Torque relationships within an analog indicating instrument Moving system Σ + − − Mechanical dynamics Jθ̈ = net torque Deflecting torque T_def θ, ω = θ̇ Control: T_c(θ) Damping: T_damp(ω) Both opposing torques return to the summing element inside the moving system.
Torque relationships within an analog indicating instrument
Figure: Torque relationships within an analog indicating instrument

Figure: Deflecting torque drives the moving system, while control and damping torques oppose displacement and motion, respectively.

If J is the moment of inertia in kg·m², the rotational equation is

J(d²θ/dt²) = Tdef − Tc − Tdamp − Tf

where θ is angular displacement in radians and Tf represents friction or other disturbance torque in N·m.

For spring control and viscous damping,

Tc = kθ, Tdamp = D(dθ/dt)

Here k is the spring constant in N·m/rad and D is the damping coefficient in N·m·s/rad. The linearized equation becomes

Jθ̈ + Dθ̇ + kθ = Tdef

For a constant input and negligible friction, the final deflection satisfies kθ = Tdef. The undamped natural angular frequency and damping ratio are

ωn = √(k/J), ζ = D/(2√(Jk))

These expressions apply when the instrument can be represented by a linear second-order rotational system. Critical damping occurs at ζ = 1. Practical instruments are often slightly underdamped to obtain a rapid but controlled response.

Control methods

Spring control uses spiral phosphor-bronze springs that also may carry current to a moving coil. It permits operation in any orientation and normally produces Tc = kθ.

Gravity control uses an adjustable weight. Its restoring torque is approximately Tc = mgr sin θ, where m is mass, g gravitational acceleration, and r the effective lever arm. It is orientation-dependent and gives a nonlinear scale.

Damping methods

  • Air-friction damping: a vane or piston moves through an enclosed air chamber.
  • Fluid-friction damping: offers strong damping but is sensitive to leakage and viscosity changes.
  • Eddy-current damping: a conducting former or disc moves through a magnetic field, producing currents whose torque opposes motion according to Lenz's law.

Damping affects settling behavior but ideally does not affect the final steady indication because damping torque vanishes when the moving system stops.

2. Energy and Co-energy Basis of Torque

Electromechanical torque can be derived by virtual work. For a magnetically linear, singly excited system carrying current i, the magnetic co-energy is

W′m(i,θ) = ½L(θ)i²

At constant current, the electromagnetic torque is

Tdef = (∂W′m/∂θ)i = ½i²(dL/dθ)

For multiple windings, mutual inductance terms must also be included. This method assumes quasi-static magnetic behavior and negligible mechanical displacement during the virtual variation.

For a voltage-excited electrostatic system with capacitance C(θ), the electrical co-energy at fixed voltage is W′e = ½C(θ)V². Therefore,

Tdef = ½V²(dC/dθ)

These relations show that many analog mechanisms move toward a position of greater inductance or capacitance.

3. Permanent-Magnet Moving-Coil Instruments

3.1 Construction and operation

A PMMC instrument has a light rectangular coil wound on an aluminium former and placed in the radial field of a permanent magnet. A soft-iron cylindrical core reduces reluctance and helps establish a nearly uniform radial flux density. Current enters and leaves the moving coil through control springs.

Correct force and magnetic-field directions in a PMMC instrument N S soft-iron core moving coil Magnetic flux density B is directed from N, through the air gaps, toward S. current into page conventional × current out of page conventional • force downward force upward
Correct force and magnetic-field directions in a PMMC instrument
Figure: Correct force and magnetic-field directions in a PMMC instrument

Figure: With flux directed from N to S, current into the page on the left experiences downward force, while current out of the page on the right experiences upward force; the two tangential forces form a couple.

The force on an active conductor of length l carrying current I in flux density B is

F = BIl sin α

where α is the angle between the current direction and magnetic field. In a PMMC radial field, α = 90°, so F = BIl.

For a rectangular coil of N turns and area A, the two forces produce

Tdef = NBAI

where B is in tesla, A in m², and I in amperes. With spring control,

θ = (NBA/k)I

Thus, a PMMC instrument has a uniform scale when the field, geometry, and spring constant remain effectively constant.

3.2 AC limitation

On direct AC connection, both current and torque reverse each half-cycle. The average torque is therefore approximately zero, and the pointer cannot indicate ordinary AC. A rectifier-PMMC instrument can measure AC, but it responds to the rectified average and is usually calibrated to indicate RMS only for a specified waveform, commonly a sine wave.

Worked Example 1: PMMC deflection

A 100-turn coil has area 4.0 × 10-4 m² in a radial field of 0.20 T. The spring constant is 8.0 × 10-6 N·m/rad. Find the deflection at 1.0 mA.

Tdef = 100(0.20)(4.0 × 10-4)(1.0 × 10-3) = 8.0 × 10-6 N·m
θ = Tdef/k = (8.0 × 10-6)/(8.0 × 10-6) = 1.00 rad

Hence the deflection is 1.00 rad, or approximately 57.3°.

3.3 Range extension

For a movement resistance Rm and full-scale current Im, a shunt Rsh extends its ammeter range to I. Equal parallel voltages give

Rsh = ImRm/(I − Im)

A series multiplier extends the movement to a voltage range V:

Rs = V/Im − Rm

These formulas assume the movement reaches full scale at Im and that connecting-lead and contact resistances are either negligible or included explicitly.

Worked Example 2: Ammeter and voltmeter ranges

A 1.0 mA, 100 Ω PMMC movement is to measure 10 A.

Rsh = (0.001)(100)/(10 − 0.001) = 0.010001 Ω ≈ 10.0 mΩ

For a 100 V voltmeter range,

Rs = 100/0.001 − 100 = 99,900 Ω

The shunt must safely carry almost 10 A, and its temperature coefficient must be low. The multiplier must withstand its power dissipation and working voltage.

4. Moving-Iron Instruments

Moving-iron instruments operate by magnetizing soft iron using a fixed coil. They may be:

  • Attraction type: a movable iron piece is drawn into a stronger magnetic field.
  • Repulsion type: similarly magnetized iron vanes repel each other.

For a magnetically linear approximation,

Tdef = ½i²(dL/dθ)

Because torque depends on i², its direction is unchanged when current reverses. Moving-iron instruments can therefore measure both AC and DC.

For AC, if dL/dθ is effectively constant over an electrical cycle,

Tavg = ½Irms²(dL/dθ)

With spring control,

θ = [Irms²/(2k)](dL/dθ)

The relationship is generally nonlinear because dL/dθ also varies with position. Scale shaping is achieved through vane geometry, but the lower scale remains compressed.

Worked Example 3: Moving-iron torque

At a particular pointer position, dL/dθ = 0.020 H/rad. For a 5.0 A RMS current,

Tavg = ½(5.0)²(0.020) = 0.250 N·m

If k = 0.50 N·m/rad and the stated derivative remains applicable at equilibrium, θ = 0.500 rad.

Moving-iron errors include magnetic hysteresis on DC, frequency-dependent coil impedance on AC, eddy currents, stray magnetic fields, friction, and iron nonlinearity.

5. Electrodynamometer Instruments

An electrodynamometer has fixed coils and a moving coil, both air-cored. Their magnetic fields interact without permanent magnets or iron in the active magnetic circuit.

For two coupled coils,

W′m = ½L1i1² + ½L2i2² + M(θ)i1i2

If the self-inductances are effectively independent of angle,

Tdef = i1i2(dM/dθ)

where M is mutual inductance in henries.

In a wattmeter, the fixed current coil carries load current i, while the moving pressure coil carries approximately v/Rp, provided its inductive reactance is negligible relative to pressure-circuit resistance Rp. The average torque is then

Tavg = (dM/dθ)(1/Rp) average[vi]

For sinusoidal voltage and current,

average[vi] = VI cos φ = P

Thus the wattmeter indicates true active power under its specified frequency and waveform conditions. Pressure-coil inductance introduces phase-angle error, especially at low power factor. Compensation resistors and capacitors may be used.

Electrodynamometer instruments operate on AC and DC and can serve as transfer standards, but they have relatively weak torque, higher cost, sensitivity to stray fields, and appreciable power consumption.

6. Induction Instruments

Induction instruments require alternating magnetic flux and therefore do not operate on DC. Two alternating fluxes displaced in time and space induce eddy currents in an aluminium disc or drum. Their interaction produces torque.

A useful design relation is

Tavg = KΦ1Φ2 sin δ

where Φ1 and Φ2 are flux amplitudes, δ is their phase displacement, and K depends on geometry, frequency, and rotor electrical properties. This expression is not a universal constant-parameter law; it is a practical relation for a specified induction mechanism and operating range.

Induction energy meter

In a single-phase energy meter, the shunt magnet produces flux approximately proportional to voltage and arranged nearly 90° out of phase with it. The series magnet produces flux approximately proportional to load current. The average driving torque is designed to satisfy

Tdrive ∝ VI cos φ

A permanent magnet produces a braking torque approximately proportional to disc speed n:

Tbrake ∝ n

At steady speed, driving and braking torques balance, so speed is proportional to power. Total revolutions are therefore proportional to energy:

E = N/Kh

where N is disc revolutions and Kh is the meter constant in rev/kWh. Alternatively, some meters specify its reciprocal in Wh/rev; the convention must always be checked.

Errors arise from friction, frequency variation, voltage variation, braking-magnet temperature, phase adjustment, creeping at no load, harmonics, and mechanical wear.

7. Electrostatic Instruments

Electrostatic instruments use the force between charged conductors. Their moving system changes the capacitance as it rotates. At fixed applied voltage,

Tdef = ½V²(dC/dθ)

Because torque is proportional to voltage squared, the mechanism responds to both AC and DC. For AC, if capacitance geometry is effectively constant during each electrical cycle,

Tavg = ½Vrms²(dC/dθ)

Electrostatic voltmeters draw only the small current needed to charge their capacitance and therefore offer exceptionally high input impedance. They are well suited to high-voltage measurement and can operate over a broad frequency range, subject to insulation, stray capacitance, dielectric loss, and resonance constraints.

Worked Example 4: Electrostatic torque

If dC/dθ = 100 pF/rad at 5.0 kV,

Tdef = ½(5000)²(100 × 10-12) = 1.25 × 10-3 N·m

The available torque is only 1.25 mN·m, illustrating why electrostatic instruments need light moving systems and careful shielding.

8. Thermal Instruments

Thermal instruments measure the heating effect of current. In a hot-wire instrument, expansion of a heated wire moves the pointer. In a thermocouple instrument, the measured current heats an element, and a thermocouple converts its temperature rise into a DC voltage that drives a PMMC movement.

A simplified thermal model is

H(d∆T/dt) + G∆T = i²R

where H is thermal capacitance in J/K, G is thermal conductance in W/K, ∆T is temperature rise in K, and R is heater resistance in ohms. At steady state,

∆T = Irms²R/G

This result assumes approximately constant resistance and that the thermal element averages the electrical heating over time. Thermal instruments provide true-RMS response within their rated crest factor, frequency, and thermal limits. They are valuable for radio-frequency measurements because their principle does not depend on magnetic response at the signal frequency.

Limitations include slow response, ambient-temperature sensitivity, heater overload risk, thermocouple aging, low output, and a nonlinear scale.

9. Range Extension with Instrument Transformers

High AC currents and voltages are commonly measured using current transformers (CTs) and voltage or potential transformers (VTs/PTs).

For an ideal CT,

Ip/Is = Ns/Np

For an ideal VT,

Vp/Vs = Np/Ns

Actual transformers exhibit ratio and phase-angle errors caused by magnetizing current, winding impedance, and burden. The burden includes the instrument, leads, and connections and must not exceed the rated value.

A CT secondary must never be open-circuited while primary current flows. With no secondary ampere-turns to oppose the primary magnetomotive force, core flux and secondary voltage can become dangerously high. A CT secondary should be shorted through an approved terminal before an instrument is disconnected. By contrast, a VT secondary must not be short-circuited.

10. Error Sources and Performance Evaluation

An indication may be represented as

measured value = true value + systematic error + random variation

Important analog-instrument errors include:

Error sourceConsequenceTypical control
Friction and stictionDead zone and poor repeatabilityJewel bearings, taut-band suspension
TemperatureChanges coil resistance, springs, magnetsManganin resistors, compensation, calibration
HysteresisDifferent upscale and downscale readingsLow-hysteresis materials
Stray fieldsUnwanted torqueMagnetic shielding, careful location
Frequency variationChanges coil impedance and phaseFrequency-rated calibration
Waveform distortionError in average-responding instrumentsTrue-RMS thermal method or waveform correction
ParallaxIncorrect scale readingMirror scale and perpendicular viewing
LoadingAlters the measured circuitHigh voltmeter resistance; low ammeter resistance
Aging and wearCalibration driftPeriodic verification and maintenance

Loading

A voltmeter with resistance Rv placed across a component is electrically in parallel with it. A finite Rv reduces the equivalent resistance and can lower the measured voltage. An ammeter introduces its resistance in series, producing a burden voltage Vb = IRa. Instrument selection must consider both nominal accuracy and disturbance of the measured system.

Accuracy classes are commonly expressed as a percentage of full-scale value rather than a percentage of reading. Consequently, relative error becomes large near the bottom of the scale. An appropriate range should place normal readings well above the scale minimum without risking overload.

11. Comparative Selection

InstrumentInput capabilityScaleMain strengthsMain limitations
PMMCDC; rectified ACUniformSensitive, accurate, low powerDirect AC torque averages to zero
Moving ironAC and DCNonlinearRugged, inexpensive, overload tolerantHysteresis and frequency errors
ElectrodynamometerAC and DCDepends on useMeasures true power; transfer useWeak torque and costly construction
InductionAC onlyApplication-dependentHigh operating torque; energy integrationFrequency and temperature dependence
ElectrostaticAC and DC voltageGenerally nonlinearVery high input impedance; high voltageLow torque and stray-capacitance effects
ThermalAC and DCNonlinearTrue-RMS heating response; high frequencySlow and susceptible to overload

12. Illustrative Industrial Case: Retrofitting a Motor Test Panel

Background

An illustrative manufacturing laboratory operates a 400 V, three-phase induction-motor test panel. The existing panel contains moving-iron ammeters and voltmeters, an electrodynamometer wattmeter, and an induction energy meter. A variable-frequency drive has recently been added.

Problem

The moving-iron current indication differs from a portable average-responding clamp meter, while the wattmeter readings become unstable at low motor power factor. Technicians also propose opening a CT secondary to replace an ammeter without interrupting the test.

Analysis

The drive output contains harmonic and switching components. The moving-iron instrument responds broadly to RMS current but may have frequency and waveform limitations. The average-responding clamp meter is calibrated for a sine wave and can be substantially inaccurate on distorted current. The electrodynamometer wattmeter measures average instantaneous power in principle, but pressure-coil phase error, bandwidth, and CT phase error become significant at low power factor and nonsinusoidal operation.

Opening an energized CT secondary would create a severe shock and insulation hazard. Furthermore, the induction energy meter is designed for rated power frequency and is unsuitable for arbitrary variable-frequency drive output.

Approach and solution

The laboratory retains analog moving-iron indicators for approximate visual monitoring on the sinusoidal supply side. It installs a properly rated power analyzer for drive-output measurements and verifies its bandwidth and crest-factor rating. CT test switches provide make-before-break shorting contacts. The electrodynamometer wattmeter is retained for periodic sinusoidal-frequency demonstrations and calibration exercises, not as the sole reference under distorted waveforms.

All instruments are calibrated through a traceable laboratory. Panel labels specify ratios, burdens, accuracy classes, frequency ranges, and safe isolation procedures.

Outcome

Disagreement between instruments is explained as a difference in measurement principle and bandwidth rather than treated as random malfunction. Measurement uncertainty decreases, and hazardous CT practices are eliminated.

Learning points

  • A displayed number is meaningful only within the instrument's waveform and frequency specification.
  • True-RMS current and active power are different measurement requirements.
  • Instrument transformers introduce ratio, phase, burden, and safety considerations.
  • Analog indications remain useful when supplemented by correctly specified modern reference instruments.

13. Professional Practice and Contemporary Developments

Calibration should be traceable to recognized national or international standards. The calibration record should state environmental conditions, range, uncertainty, adjustment status, and as-found and as-left results. Engineers must not report more resolution than the scale and uncertainty justify.

Relevant professional considerations include electrical installation requirements, insulation category, working voltage, protective devices, electromagnetic compatibility, and safe handling of instrument transformers. Environmental governance also requires responsible disposal of obsolete instruments containing hazardous oils, mercury switches, lead-based materials, or degraded insulation.

Contemporary analog indicators increasingly use electronic transducers and servo drives rather than relying entirely on the measured quantity to produce torque. Digital image processing can also read legacy pointer instruments remotely. Such systems may automate inspection, but calibration, perspective correction, pointer detection, uncertainty assessment, and cybersecurity remain human engineering responsibilities.

Chapter Summary

Analog electromechanical instruments balance a measurand-dependent deflecting torque against a controlling torque, while damping governs transient response. PMMC mechanisms produce torque proportional to DC current and have uniform scales. Moving-iron instruments produce torque approximately proportional to current squared and operate on AC or DC. Electrodynamometer instruments use mutual-field interaction and are especially important for power measurement. Induction instruments use alternating fluxes and induced currents, notably in energy meters. Electrostatic instruments develop voltage-squared torque with very high input impedance, while thermal instruments measure RMS heating and can operate at high frequencies.

Energy and co-energy methods provide systematic torque derivations. Practical measurement also requires correct range extension, control of loading and environmental errors, suitable waveform and frequency response, safe CT and VT procedures, traceable calibration, and informed instrument selection.

Key Terms

Deflecting torque: Torque produced by the electrical measurand.

Controlling torque: Restoring torque that establishes the final pointer position.

Damping torque: Motion-dependent torque that reduces oscillation and settling time.

PMMC instrument: Permanent-magnet moving-coil instrument primarily used for DC measurement.

Moving-iron instrument: Instrument using attraction or repulsion of magnetized soft iron.

Electrodynamometer: Air-cored instrument using interaction between fixed- and moving-coil fields.

Induction instrument: AC instrument using alternating fluxes and rotor eddy currents.

Electrostatic instrument: Instrument using force associated with electric-field energy.

Thermal instrument: Instrument that measures the heating effect of current.

Co-energy: Energy function used conveniently to calculate force or torque at constant current or voltage.

Shunt: Low resistance connected in parallel with an ammeter movement to extend current range.

Multiplier: High resistance connected in series with a voltmeter movement to extend voltage range.

Instrument transformer: Transformer that scales current or voltage and may provide isolation.

Burden: Impedance and volt-ampere load connected to an instrument-transformer secondary.

Hysteresis error: Difference caused by dependence on previous magnetic or mechanical state.

Parallax: Reading error caused by viewing the pointer and scale from an oblique angle.

Loading effect: Change in the measured circuit caused by connecting an instrument.

Accuracy class: Specified error limit, often expressed as a percentage of full scale.

True RMS: Effective value determined from the mean of the squared waveform rather than from sine-wave calibration assumptions.

Review Questions

Short Answer Questions

  1. Distinguish among deflecting, controlling, and damping torque.
  2. Why does a direct-connected PMMC instrument not indicate ordinary AC?
  3. State two important sources of error in a moving-iron instrument.
  4. Why must an energized CT secondary never be open-circuited?
  5. What is meant by the loading effect of a voltmeter?

Long Answer Questions

  1. Derive the torque and deflection equations of a PMMC instrument and explain why its scale is uniform.
  2. Derive the moving-iron torque equation using magnetic co-energy and discuss its AC operation.
  3. Explain the construction and wattmeter operation of an electrodynamometer instrument, including phase-angle error.
  4. Describe the operating principle of a single-phase induction energy meter and explain its principal adjustments and errors.
  5. Compare PMMC, moving-iron, electrodynamometer, induction, electrostatic, and thermal instruments with respect to principle, input type, scale, errors, and applications.

Analytical / Application Questions

  1. A 2 mA, 50 Ω PMMC movement must provide 5 A and 250 V ranges. Calculate the required shunt and series multiplier and determine their minimum ideal power dissipations at full scale.
  2. Develop the step-response equation for a spring-controlled instrument with viscous damping and explain how increasing the damping ratio changes overshoot and settling behavior.
  3. Select suitable instruments for measuring DC current, distorted high-frequency current, 20 kV AC voltage, sinusoidal active power, and AC energy. Justify every selection using operating principle and limitations.

Multiple Choice Questions with Answers

Question 1

Which torque ideally becomes zero when the pointer comes to rest at its final indication?

A. Deflecting torque

B. Controlling torque

C. Damping torque

D. Spring torque

Correct Answer: C. Damping torque

Question 2

A PMMC coil has 50 turns, area 2 × 10-4 m², flux density 0.10 T, and current 2 mA. What is its deflecting torque?

A. 2 × 10-7 N·m

B. 2 × 10-6 N·m

C. 2 × 10-5 N·m

D. 2 × 10-4 N·m

Correct Answer: B. 2 × 10-6 N·m

Question 3

Why can a moving-iron instrument measure both AC and DC?

A. Its torque is approximately proportional to current squared

B. It contains a permanent rectifier

C. Its damping torque reverses every half-cycle

D. Its coil carries no current

Correct Answer: A. Its torque is approximately proportional to current squared

Question 4

For an ideal spring-controlled PMMC instrument, pointer deflection is directly proportional to:

A. Current squared

B. Voltage squared only

C. Current

D. Frequency

Correct Answer: C. Current

Question 5

A 1 mA, 100 Ω PMMC movement is extended to 100 V. What series multiplier is required?

A. 9.9 kΩ

B. 10.0 kΩ

C. 99.9 kΩ

D. 100.1 kΩ

Correct Answer: C. 99.9 kΩ

Question 6

In an electrodynamometer wattmeter with negligible pressure-coil phase error, average torque is proportional to:

A. Reactive power

B. Apparent power only

C. Average instantaneous power

D. Current frequency only

Correct Answer: C. Average instantaneous power

Question 7

Which instrument fundamentally requires alternating magnetic flux for operation?

A. PMMC instrument

B. Induction instrument

C. Electrostatic instrument

D. Hot-wire instrument

Correct Answer: B. Induction instrument

Question 8

If dC/dθ = 80 pF/rad and an electrostatic instrument is energized at 2 kV, what is its torque?

A. 0.016 mN·m

B. 0.16 mN·m

C. 1.6 mN·m

D. 16 mN·m

Correct Answer: B. 0.16 mN·m

Question 9

What is the safest required action before disconnecting an ammeter from an energized CT secondary circuit?

A. Open the CT secondary

B. Short the CT primary

C. Short the CT secondary through an approved test terminal

D. Connect the CT secondary to a capacitor

Correct Answer: C. Short the CT secondary through an approved test terminal

Question 10

Which instrument is most directly suited to true-RMS current measurement at radio frequencies within its specified limits?

A. Thermocouple instrument

B. PMMC instrument without a rectifier

C. Induction energy meter

D. Gravity-controlled moving-iron instrument

Correct Answer: A. Thermocouple instrument

Electrical Measurements & Instrumentation
Chapter 3 of 14