Electrical Measurement & Instrumentation

Chapter 5
Electronic Instruments and Time–Frequency Measurement

Post Graduate Program in Electrical Engineering
Electrical Measurement & Instrumentation   •   Post Graduate

Electronic Instruments and Time–Frequency Measurement

Introduction

Electronic instruments convert electrical quantities into forms suitable for amplification, digitization, processing, display and storage. Unlike purely electromechanical instruments, they can achieve high input impedance, wide bandwidth, automated ranging, waveform capture and statistical analysis. Their accuracy nevertheless depends on analogue front ends, reference standards, sampling processes, triggering and calibration—not merely on display resolution.

Time and frequency measurement is especially important because frequency can be compared with exceptionally stable references. Applications include communication systems, power-electronic converters, protection relays, radar, navigation, semiconductor testing, synchronized grids and precision metrology. At postgraduate level, instrument readings must be interpreted through measurement models, uncertainty, bandwidth and noise rather than treated as exact values.

Learning Objectives

After studying this chapter, the reader should be able to:

  • explain the architecture and limitations of major electronic instruments;
  • distinguish resolution, accuracy, precision, bandwidth and sensitivity;
  • analyse integrating voltmeters, digital oscilloscopes and spectrum analysers;
  • derive direct and reciprocal frequency-measurement equations;
  • evaluate time-interval and phase measurements, including channel-delay errors;
  • quantify phase noise and its relationship to timing jitter;
  • construct uncertainty budgets for time–frequency measurements;
  • select suitable instruments, probes, references and measurement procedures.

1. Measurement Architecture of Electronic Instruments

A general electronic instrument consists of an input interface, signal-conditioning network, conversion subsystem, processor, reference source and output interface.

Functional architecture of an electronic measuring instrument Input andprotection Signalconditioning ADC orcomparator Digitalprocessing Display, storageand interface Reference andtimebase
Functional architecture of an electronic measuring instrument
Figure: Functional architecture of an electronic measuring instrument

Figure: Functional architecture of a modern electronic measuring instrument.

1.1 Input loading and transfer behaviour

An instrument alters the circuit to which it is connected. For a source represented by voltage Vs and impedance Zs, connected to an instrument with input impedance Zi, the measured input is

Vi = Vs Zi/(Zs + Zi)

The expression applies to linear impedances at the frequency of interest. Negligible loading requires |Zi| ≫ |Zs|, while phase accuracy also requires consideration of the complex impedance angle. An oscilloscope probe, for example, may present approximately 10 MΩ at low frequency but much lower impedance at high frequency because of input capacitance.

For a linear time-invariant instrument, the frequency-domain response is

Y(jω) = H(jω)X(jω)

where H(jω) includes magnitude and phase response. A flat magnitude response alone does not guarantee faithful transient reproduction; phase linearity and group delay are also important.

1.2 Instrument performance terms

TermMeaning
ResolutionSmallest displayed or distinguishable increment
AccuracyCloseness to the accepted reference value, normally expressed through uncertainty or limits
PrecisionCloseness among repeated observations
SensitivityChange in output per unit change in input
LinearityCloseness of the input–output relation to a specified straight line
BandwidthFrequency range over which specified performance is maintained
Dynamic rangeRatio between the largest usable signal and the noise or distortion floor
DriftSlow change in response with time or environmental conditions

A six-digit display does not imply six-digit accuracy. Display resolution, calibration uncertainty, noise, temperature coefficient and long-term drift must be considered separately.

2. Digital Voltmeters and Multimeters

2.1 ADC-based measurement

A digital voltmeter normally includes protection, attenuation, amplification, filtering, analogue-to-digital conversion and numerical correction. If an ideal n-bit bipolar or unipolar converter spans a total input range VR, its code width is

q = VR/2n

For ideal rounding to the nearest code, quantization error lies between -q/2 and +q/2. If uniformly distributed, its standard uncertainty is

uq = q/√12

This model excludes offset, gain error, nonlinearity and noise. Real instruments frequently exhibit code transition noise exceeding ideal quantization.

2.2 Dual-slope integrating conversion

Dual-slope conversion is widely used in precision digital multimeters. The unknown voltage is integrated for a fixed interval T1. A reference voltage of opposite polarity is then applied until the integrator returns to zero. For an ideal integrator and constant input,

Vx T1 = Vref T2

and therefore

Vx = Vref T2/T1

Here Vx and Vref denote voltage magnitudes, and T2 is the deintegration interval. The integrating capacitor and resistor cancel from the ratio, improving stability. Choosing T1 as an integer multiple of the mains period strongly rejects 50 or 60 Hz interference. Rapidly varying signals, however, are averaged rather than captured instantaneously.

2.3 AC measurements

AC voltmeters may use average-responding, peak-responding or true-RMS conversion. The RMS value over observation time T is

Vrms = √[(1/T)∫0T v²(t)dt]

A true-RMS instrument must still satisfy specified bandwidth and crest-factor limits. A high-crest-factor waveform can overload the input stage even when its RMS value is within range. The user should therefore verify peak voltage, bandwidth and permissible crest factor.

3. Digital Oscilloscopes

A digital storage oscilloscope samples the input, stores digital records and reconstructs a displayed waveform. Its main subsystems are the vertical attenuator and amplifier, sample-and-hold circuit, ADC, acquisition memory, trigger system and timebase.

3.1 Sampling and aliasing

For a signal strictly band-limited to B, uniform sampling permits ideal reconstruction when

fs > 2B

where fs is sampling frequency. This is a theoretical condition requiring an ideal reconstruction process. Practical oscilloscopes use sampling rates substantially above twice the analogue bandwidth and incorporate anti-aliasing or bandwidth-limiting filters.

Signals above the Nyquist frequency may appear at false frequencies. For a sinusoidal component of frequency f, aliases occur at

fa = |f - kfs|

where integer k is chosen so that fa lies in the displayed baseband. Equivalent-time sampling can provide high apparent sample rates for repetitive, stable waveforms but cannot faithfully acquire a unique single-shot transient.

3.2 Bandwidth and rise time

For a single-pole response, the approximate relationship between 10–90% rise time and −3 dB bandwidth is

tr ≈ 0.35/BW

This coefficient is not universal; it changes with filter response. For cascaded approximately Gaussian subsystems, a useful estimate is

tr,meas ≈ √(tr,signal² + tr,scope² + tr,probe²)

The expression is an approximation and should not be applied indiscriminately to strongly non-Gaussian responses.

3.3 Triggering and timing extraction

Trigger systems establish the record’s temporal reference. Trigger uncertainty arises from input noise, finite comparator slope sensitivity and internal timebase uncertainty. For a threshold crossing with local slew rate S = dv/dt and RMS voltage noise σv, the small-noise timing jitter is approximately

σt ≈ σv/|S|

Thus, steeper edges and lower noise improve timing precision. Hysteresis can reduce false triggers but may introduce level-dependent timing shifts.

4. Spectrum and Signal Analysis

A swept-tuned spectrum analyser mixes the input with a local oscillator and measures energy passing through an intermediate-frequency resolution bandwidth. A fast-Fourier-transform analyser digitizes a time record and computes its spectrum.

Key settings include resolution bandwidth (RBW), video bandwidth, span, detector type, window function and averaging. Reducing RBW separates closely spaced components and lowers displayed noise power, but increases sweep or record duration. Windowing controls spectral leakage caused by finite observation time. Amplitude accuracy also depends on impedance matching, attenuator accuracy, mixer compression and detector calibration.

Noise spectral density is commonly reported in V²/Hz, A²/Hz, W/Hz, dBm/Hz or dBc/Hz, depending on context. A noise density cannot be compared with integrated noise until the effective noise bandwidth is included.

5. Frequency Measurement

5.1 Direct frequency counting

In direct counting, the instrument counts input cycles during a gate interval Tg. If Nx cycles are counted,

f̂x = Nx/Tg

where f̂x is the frequency estimate in hertz. If the gate is asynchronous to the input and only complete cycles are counted, endpoint phase can cause approximately ±1-count error. Modelling this bounded count error as rectangular gives a standard count uncertainty of 1/√3 count.

Some instruments instead determine fractional-cycle information or round an interpolated count to the nearest cycle. Only for a genuine nearest-count rounding model is the error bound ±0.5 count and the corresponding standard uncertainty 1/√12 count. The convention must therefore be stated rather than mixing these two models.

For the asynchronous ±1-count model, the relative quantization bound is approximately

|δfq|/fx ≤ 1/Nx = 1/(fxTg)

Longer gates improve resolution but slow updates and may average frequency variation.

Worked Example 1: Direct counting

A nominal 10 MHz source is counted for 0.1 s, producing 1,000,000 counts. The estimate is 10 MHz. Under the asynchronous ±1-count model, the relative error bound is

1/Nx = 1/1,000,000 = 1 × 10-6

The standard relative uncertainty from rectangular count ambiguity is 1/(√3Nx) = 5.77 × 10-7. If the independent reference-timebase standard uncertainty is 2 × 10-8, the combined relative standard uncertainty is

ur = √[(5.77 × 10-7)² + (2 × 10-8)²] ≈ 5.78 × 10-7

The count ambiguity dominates.

5.2 Reciprocal frequency counting

A reciprocal counter measures the elapsed time for an integer number K of input periods. Let C be the number of reference-clock ticks accumulated and tc the reference-clock period. Neglecting interpolation and systematic delays,

T̂K = Ctc

and

f̂x = K/T̂K = K/(Ctc)

The input-cycle count K and reference-clock count C are distinct quantities. If the elapsed-time quantization is approximately one clock tick, the relative resolution is approximately 1/C. Reciprocal counting is particularly advantageous at low input frequencies because timing many input periods yields a large reference count.

Worked Example 2: Reciprocal measurement

A 1 kHz signal is measured over K = 100 periods using a 100 MHz clock, for which tc = 10 ns. The expected elapsed time is 0.1 s and

C = 0.1/(10 × 10-9) = 10,000,000

Hence

f̂x = 100/[10,000,000(10 × 10-9)] = 1000 Hz

A one-tick timing uncertainty corresponds to an approximate relative resolution of 10-7, before timebase and trigger uncertainties are included.

5.3 Heterodyne and comparison methods

Very high frequencies may be prescaled or mixed with a known local oscillator. If a mixer generates a difference frequency,

fb = |fx - fLO|

The unknown frequency is determined only after the correct sum or difference branch and harmonic order are identified. Frequency ratio counters compare two sources directly and can suppress uncertainty associated with a shared timebase.

6. Time-Interval Measurement

A time-interval counter estimates the separation between start event A and stop event B. Let the true event times be tA and tB, so that

Δt = tB - tA

Suppose channel A has fixed delay bA and random timestamp error εA, while channel B has corresponding quantities bB and εB. Including quantization or interpolation error εq, the measured interval is

Δt̂ = Δt + bB - bA + εB - εA + εq

The sign follows directly from subtracting the measured A timestamp from the measured B timestamp. Calibration must therefore determine the differential channel delay bB-bA, not merely the individual cable lengths.

Worked Example 3: Channel-delay correction

The true interval is 250 ns. Channel delays are bA = 1.2 ns and bB = 2.0 ns. With negligible random error,

Δt̂ = 250 + 2.0 - 1.2 = 250.8 ns

The corrected result is obtained by subtracting the measured differential delay of 0.8 ns.

Modern counters use tapped delay lines, time-to-digital converters, Vernier methods or analogue interpolation to achieve resolution finer than one reference-clock period. Fine resolution does not guarantee equal accuracy because differential nonlinearity, temperature drift and trigger walk may remain.

Trigger walk occurs when signals of different amplitudes cross a fixed threshold at different relative points. Constant-fraction triggering reduces amplitude dependence. Cable delays, connector repeatability, impedance discontinuities and threshold settings are critical in sub-nanosecond measurements.

7. Phase Measurement

For two sinusoidal signals of common frequency f, a time displacement Δt corresponds to phase difference

Δφ = 2πfΔt

where Δφ is in radians. In degrees, Δφ = 360°fΔt. The result is normally known modulo 2π, so additional information is required to identify whole-cycle ambiguity.

Phase can be measured using zero crossings, digital timestamps, mixers, lock-in amplifiers, phase detectors or cross-spectral methods. Mixer-based detectors require operation within their approximately linear phase range. Cross-correlation can suppress uncorrelated instrument noise when two independent channels observe the same source.

8. Phase Noise, Frequency Stability and Timing Jitter

8.1 Quantitative phase-noise model

A practical oscillator can be represented as

v(t) = V0[1 + a(t)]cos[2πf0t + φ(t)]

where f0 is carrier frequency, a(t) is fractional amplitude fluctuation and φ(t) is random phase fluctuation in radians.

The one-sided phase-fluctuation power spectral density S_φ(fm) has units rad²/Hz, where fm is offset frequency from the carrier. Under the small-phase-noise approximation with symmetrical sidebands, single-sideband phase noise is

L(fm) = S_φ(fm)/2

in linear units. It is commonly reported as

LdBc/Hz(fm) = 10log10[L(fm)]

The relation does not apply without qualification when amplitude noise, large modulation or asymmetric sidebands dominate.

Timing fluctuation is related to phase fluctuation by

x(t) = φ(t)/(2πf0)

and its spectral density is

Sx(fm) = S_φ(fm)/(2πf0)²

with units s²/Hz. RMS timing jitter integrated over offset frequencies f1 to f2 is

σt = [1/(2πf0)]√[∫f1f2S_φ(fm)dfm]

or, using linear single-sideband phase noise,

σt = [1/(2πf0)]√[2∫f1f2L(fm)dfm]

The integration limits and removal of discrete spurs must be stated because jitter depends strongly on measurement bandwidth.

Worked Example 4: Phase noise to jitter

A 100 MHz oscillator has flat phase noise of −120 dBc/Hz between 1 kHz and 100 kHz. Assume negligible noise elsewhere in the stated jitter bandwidth and no discrete spurs. The linear single-sideband density is L = 10-12/Hz, so

∫S_φdf = 2(10-12)(100,000 - 1,000) = 1.98 × 10-7 rad²

Therefore,

σt = √(1.98 × 10-7)/[2π(100 × 106)] ≈ 7.08 × 10-13 s

The integrated RMS jitter is approximately 0.708 ps.

8.2 Allan deviation

Ordinary standard deviation does not adequately characterize many oscillator noise processes. Let ȳk be the average fractional frequency over the kth non-overlapping interval of duration τ. The two-sample Allan deviation is

σy(τ) = √{(1/2)⟨[ȳk+1 - ȳk]²⟩}

It indicates stability as a function of averaging time and helps distinguish white phase noise, white frequency noise, flicker processes and random walk. Modified Allan deviation provides additional discrimination and is sensitive to measurement bandwidth and sampling strategy.

9. Reference Oscillators and Traceability

Frequency counters and oscilloscopes inherit the characteristics of their timebases. Common references include temperature-compensated crystal oscillators, oven-controlled crystal oscillators, rubidium standards and GNSS-disciplined oscillators.

Important attributes are initial offset, ageing, temperature sensitivity, short-term noise, warm-up behaviour and holdover performance. GNSS discipline provides long-term traceability but can be affected by antenna faults, interference or spoofing. A local oscillator may therefore be combined with monitoring and holdover algorithms.

Metrological traceability requires an unbroken, documented chain of calibrations to recognized standards, with uncertainty stated at each link. Calibration adjusts or characterizes an instrument; it does not make the instrument permanently error-free.

10. Measurement Uncertainty

For a result y = g(x1,x2,…,xn), the first-order combined standard uncertainty is

uc²(y) = ∑i=1n(ci ui)² + 2∑i<jcicj u(xi,xj)

where ci = ∂g/∂xi is a sensitivity coefficient, ui is a standard uncertainty and u(xi,xj) is covariance. The simplified root-sum-square form is valid only when inputs are uncorrelated.

A time–frequency uncertainty budget may include:

  • reference calibration and ageing;
  • counter quantization or interpolation;
  • trigger noise and threshold walk;
  • channel-delay calibration;
  • temperature effects;
  • cable and connector repeatability;
  • source instability during observation;
  • statistical estimation uncertainty.

Expanded uncertainty is commonly reported as U = kuc, where coverage factor k is selected for the required coverage and effective degrees of freedom. Reporting should state the measurement model, result, uncertainty, coverage factor, conditions and traceability.

11. Instrument Selection and Best Practice

  1. Define the measurand, bandwidth and required uncertainty before selecting equipment.
  2. Distinguish display resolution from validated measurement accuracy.
  3. Use probes and cables with appropriate impedance, voltage rating and bandwidth.
  4. Allow references and analogue circuits to warm up and reach thermal equilibrium.
  5. Share a common reference when comparing phase-coherent instruments, while checking ground-loop risks.
  6. Characterize differential channel delay using the actual cables and threshold settings.
  7. Prevent aliasing by adequate sampling and filtering.
  8. Record gate time, averaging, RBW, trigger level and interpolation settings.
  9. Check instrument self-test and calibration status.
  10. Preserve raw records and configuration metadata for reproducibility.

12. Illustrative Case Study: Qualification of a Converter Control Clock

Background

This is an illustrative case. A laboratory is validating a 1 kHz synchronization output generated by a grid-connected power converter. Requirements are frequency error below 2 ppm, channel-to-channel latency below 300 ns and integrated clock jitter below 1 ps over 1 kHz–100 kHz.

Problem

Direct counting with a 0.1 s gate would produce only 100 input counts, giving an asynchronous count-error bound of approximately 1%. This is inadequate for a 2 ppm requirement. The latency measurement also involves two counter channels with unequal delays.

Analysis and approach

The team uses a reciprocal counter disciplined by a calibrated 100 MHz reference. It measures 1000 input periods, corresponding to approximately 1 s and 100 million reference ticks. The one-tick relative resolution is approximately 10-8, below the required 2 ppm, although timebase uncertainty and source variation must still be included.

A common pulse is then split into both timing channels. The measured B-minus-A offset is 0.74 ns, establishing bB-bA = 0.74 ns. Subsequent converter latency results are corrected by subtracting this value.

Phase noise is measured using a cross-correlation analyser to reduce uncorrelated analyser noise. Integration is limited to the specified 1 kHz–100 kHz band, and identified deterministic spurs are reported separately rather than silently removed.

Outcome

The measured frequency is 1000.0008 Hz, corresponding to +0.8 ppm. Corrected latency is 247.6 ns. Integrated random jitter is 0.76 ps RMS, with two deterministic spurs documented separately. All three principal requirements are satisfied.

Learning points

The case demonstrates that measurement method must follow the specification. Reciprocal counting avoids poor low-frequency resolution, differential-delay calibration prevents systematic timing bias, and jitter results are meaningful only when bandwidth and spur treatment are declared.

13. Challenges, Safety and Contemporary Developments

High-speed instruments face a trade-off among bandwidth, noise, memory depth, linearity and power consumption. Automated processing can hide assumptions about filtering, interpolation and edge selection. Software-defined instruments and machine-learning algorithms increasingly assist anomaly detection, calibration scheduling and waveform classification, but they should not replace traceable measurement models. Training data, algorithm versions and automated decisions require validation.

Electrical safety must follow instrument voltage, current and measurement-category ratings. IEC 61010 principles distinguish CAT II, CAT III and CAT IV environments according to transient exposure. A probe’s voltage rating can decrease with frequency. Protective-earth removal is not an acceptable method for eliminating ground loops; differential probes, isolation methods and appropriate test architecture should be used instead.

Data integrity is also a governance issue. Raw data, correction factors and rejected observations should be auditable. GNSS-dependent timing systems require monitoring against jamming and spoofing. Results used for regulatory compliance should include calibration status, uncertainty and environmental conditions.

Chapter Summary

Electronic instruments combine analogue conditioning, conversion, digital processing and a reference system. Their performance is governed by loading, bandwidth, noise, sampling, triggering and calibration as well as nominal resolution. Integrating voltmeters provide strong interference rejection, while oscilloscopes and spectrum analysers require careful treatment of aliasing, response and measurement bandwidth.

Direct frequency counting estimates frequency from input cycles within a gate and is limited by the stated endpoint-count convention. Reciprocal counting separately counts input periods and reference-clock ticks, providing superior low-frequency resolution. Time-interval measurements require correction for the signed differential delay bB-bA. Phase noise is quantified through spectral density or dBc/Hz and can be integrated over declared offset limits to obtain timing jitter. Reliable professional measurements require traceability, uncertainty evaluation, suitable probes, reproducible settings, safety compliance and transparent data handling.

Key Terms

Accuracy: Closeness of a result to the accepted reference value, expressed through error limits or uncertainty.

Aliasing: Appearance of sampled frequency components at incorrect lower frequencies.

Allan deviation: Two-sample statistic used to characterize frequency stability versus averaging time.

Bandwidth: Frequency range over which an instrument meets specified performance.

Crest factor: Ratio of waveform peak magnitude to RMS value.

Differential channel delay: Difference between propagation delays of two measurement channels.

Direct counting: Frequency measurement by counting input cycles during a known gate interval.

Dynamic range: Ratio between the largest usable signal and the effective noise or distortion floor.

Jitter: Short-term timing variation of signal events from their ideal positions.

Phase noise: Spectral characterization of random phase fluctuations around a carrier.

Quantization: Mapping of a continuous input to discrete digital levels or time bins.

Reciprocal counting: Frequency measurement based on timing a known number of input periods with a reference clock.

Resolution bandwidth: Spectrum-analyser bandwidth that controls frequency discrimination and displayed noise.

Sensitivity coefficient: Derivative describing how an input quantity affects a measurement result.

Timebase: Reference oscillator that establishes an instrument’s timing scale.

Traceability: Documented calibration chain linking a result to recognized standards.

Trigger jitter: Random temporal variation in trigger recognition.

Trigger walk: Timing shift caused by amplitude-dependent threshold crossing.

Review Questions

Short Answer Questions

  1. Distinguish resolution, precision and accuracy in an electronic instrument.
  2. Why does a dual-slope converter reject periodic interference?
  3. State the difference between direct and reciprocal frequency counting.
  4. What is differential channel delay?
  5. Why must jitter integration limits be reported?

Long Answer Questions

  1. Explain the architecture of a digital storage oscilloscope and discuss its major error sources.
  2. Derive the dual-slope voltmeter relationship and evaluate its advantages and limitations.
  3. Compare direct counting, reciprocal counting and heterodyne frequency measurement.
  4. Explain phase-noise spectral density, single-sideband phase noise and their relationship to timing jitter.
  5. Develop an uncertainty budget for a two-channel time-interval measurement.

Analytical / Application Questions

  1. A 2 MHz source is counted for 0.5 s. Determine the asynchronous ±1-count relative error bound and corresponding rectangular standard uncertainty.
  2. A reciprocal counter times 500 periods using a 200 MHz reference and records 25,000,000 reference ticks. Calculate the measured input frequency.
  3. Two timing channels have delays of 3.6 ns and 2.1 ns for channels A and B, respectively. If the indicated B-minus-A interval is 98.0 ns, calculate the corrected interval.

Multiple Choice Questions with Answers

Question 1

In reciprocal frequency counting, K input periods produce C reference-clock ticks of duration tc. Which expression gives the estimated input frequency?

A. C/(Ktc)

B. K/(Ctc)

C. 1/(Ktc)

D. Ctc/K²

Correct Answer: B. K/(Ctc)

Question 2

An asynchronous direct counter records 200,000 cycles during its gate. Under the stated ±1-count endpoint model, what is the approximate relative error bound?

A. 2 × 10-4

B. 1 × 10-5

C. 5 × 10-6

D. 2.5 × 10-6

Correct Answer: C. 5 × 10-6

Question 3

For measured channel timestamps containing delays bA and bB, which systematic term appears in the measured B-minus-A interval?

A. bA+bB

B. bA-bB

C. -bA-bB

D. bB-bA

Correct Answer: D. bB-bA

Question 4

Under the small-phase-noise and symmetric-sideband assumptions, linear single-sideband phase noise is related to one-sided phase PSD by which expression?

A. L(fm)=S_φ(fm)/2

B. L(fm)=2S_φ(fm)

C. L(fm)=√S_φ(fm)

D. L(fm)=S_φ²(fm)

Correct Answer: A. L(fm)=S_φ(fm)/2

Question 5

A dual-slope voltmeter integrates an unknown voltage for 100 ms and requires 40 ms to return to zero using a 10 V opposite-polarity reference. What is the unknown voltage magnitude?

A. 2.5 V

B. 4.0 V

C. 10 V

D. 25 V

Correct Answer: B. 4.0 V

Question 6

A 70 MHz sinusoid is sampled at 100 MHz. Which baseband alias frequency is observed?

A. 20 MHz

B. 30 MHz

C. 50 MHz

D. 70 MHz

Correct Answer: B. 30 MHz

Question 7

For ideal rounding quantization with code width q, what is the standard uncertainty when the error is uniformly distributed over -q/2 to +q/2?

A. q/2

B. q/√3

C. q/√12

D. q/12

Correct Answer: C. q/√12

Question 8

A single-pole oscilloscope has a bandwidth of 100 MHz. Its approximate 10–90% rise time is:

A. 0.35 ns

B. 3.5 ns

C. 10 ns

D. 35 ns

Correct Answer: B. 3.5 ns

Question 9

A 1 kHz input is measured over 100 periods using a 100 MHz reference clock. Approximately how many reference ticks are accumulated?

A. 100,000

B. 1,000,000

C. 10,000,000

D. 100,000,000

Correct Answer: C. 10,000,000

Question 10

Which practice is essential when reporting RMS jitter derived from phase-noise data?

A. Report only the carrier amplitude

B. State the phase-noise integration limits

C. Replace the calibrated reference with the oscilloscope display rate

D. Express all discrete spurs as DC offset

Correct Answer: B. State the phase-noise integration limits