Chapter 4
Classical Analog Instruments and Bridge Methods
Post Graduate Program in Electrical Engineering
Electrical Measurement & Instrumentation
•
Post Graduate
Classical Analog Instruments and Bridge Methods
Introduction
Classical analog instruments convert an electrical quantity into a mechanical displacement, usually by balancing electromagnetic or electrostatic deflecting torque against a controlling torque. Bridge methods instead compare an unknown component with calibrated standards and determine its value from a null condition. The two approaches remain important because they reveal the physical basis of measurement, loading, sensitivity, uncertainty and traceability.
Although digital instruments now dominate routine measurements, analog mechanisms continue to be used in rugged indicators, educational laboratories, protection systems and reference electromechanical devices. Bridge principles also survive in resistance thermometry, strain-gauge interfaces, impedance analyzers, transformer testing and dielectric-loss measurement.
Learning Objectives
After studying this chapter, the reader should be able to:
derive the static and dynamic equations of analog indicating instruments;
compare PMMC, moving-iron, electrodynamic, induction and electrostatic instruments;
explain controlling, damping, loading and scale characteristics;
derive the general balance condition of a four-arm bridge;
apply Wheatstone, Kelvin double, Schering and Wien bridges correctly;
evaluate bridge sensitivity, parasitic effects and measurement uncertainty;
select suitable instruments and bridges for practical measurements;
apply guarding, shielding, calibration and safety practices.
1. Analog Indicating Instruments
1.1 Functional model
An analog indicating instrument contains a sensing element, a torque-producing mechanism, a moving system, a controlling system, damping and a calibrated scale. Its motion is governed by
where J is the rotational inertia in kg·m², D is the damping coefficient in N·m·s/rad, K is the spring constant in N·m/rad, \theta is angular displacement in radians and Td is deflecting torque in N·m.
For a constant input at steady state, velocity and acceleration vanish, giving
K\theta=Td
The dynamic equation is a second-order system with natural angular frequency and damping ratio
Underdamping produces oscillation, overdamping gives a slow response, and critical damping, \zeta=1, provides the fastest nonoscillatory response. Practical instruments are commonly designed to be slightly underdamped or approximately critically damped.
1.2 Controlling and damping torque
Controlling torque establishes a unique steady indication and returns the pointer to zero. Spiral springs are commonly used because they also conduct current to a moving coil. Gravity control is position-sensitive and is now uncommon.
Damping acts only while the moving system is in motion. Principal methods are:
Air-friction damping: a vane or piston moves through an enclosed air chamber.
Fluid-friction damping: effective but sensitive to temperature and orientation.
Eddy-current damping: motion of a conductor in a magnetic field induces currents opposing the motion according to Lenz's law.
Damping must not be confused with controlling torque: damping determines transient behavior, while control determines the final deflection.
2. Principal Classical Instrument Mechanisms
2.1 Permanent-magnet moving-coil instrument
A permanent-magnet moving-coil (PMMC) mechanism has a coil of N turns and effective area A in a radial magnetic field of flux density B. With coil current I, its deflecting torque is
Td=NBAI
This expression assumes a uniform radial field and that all active conductors experience the designed flux density. Equating deflecting and spring-control torques gives
\theta=\frac{NBA}{K}I
Thus PMMC instruments have a uniform scale. Their high sensitivity and low power consumption make them suitable for DC ammeters, voltmeters and galvanometers.
On sinusoidal AC, current reverses every half-cycle and so does torque. The mean torque is zero unless a rectifier is added. A rectifier-type PMMC instrument responds approximately to the average of the rectified waveform but is usually calibrated in RMS units for a sine wave. Its indication is therefore waveform-dependent.
A PMMC ammeter uses a low-resistance shunt. If a movement of resistance Rm carries full-scale current Im, while the desired range is I, the shunt resistance is
Rs=\frac{ImRm}{I-Im}
A PMMC voltmeter uses a series multiplier:
Rseries=\frac{V}{Im}-Rm
These equations assume negligible temperature change and accurate knowledge of movement resistance.
2.2 Moving-iron instrument
Moving-iron instruments use attraction or repulsion of soft iron in the magnetic field of a current-carrying coil. The magnetic co-energy is approximately
W'=\frac{1}{2}L(\theta)i2
and the instantaneous torque is
Td=\frac{1}{2}i2\frac{dL}{d\theta}
Because torque depends on i2, its average is positive for either current direction. Moving-iron instruments therefore operate on both DC and AC and indicate approximately RMS current when magnetic behavior is sufficiently linear. Their scales are generally nonlinear, particularly near zero.
Important errors include hysteresis, frequency-dependent coil impedance, eddy currents and waveform effects. Their ruggedness and overload capability remain useful in switchboards and industrial panels.
2.3 Electrodynamic instruments
An electrodynamic instrument uses fixed and moving air-cored coils. If their mutual inductance is M(\theta), the torque component of interest is
Td=i1i2\frac{dM}{d\theta}
For a wattmeter, the fixed current coil carries load current and the moving pressure coil carries a current proportional to voltage. The average torque is then proportional to active power:
Here V and I are RMS values and \phi is their phase difference for sinusoidal signals. Electrodynamic wattmeters can operate on AC and DC and are valuable as transfer standards. Their limitations include weak torque-to-weight ratio, stray-field sensitivity and pressure-coil phase error.
2.4 Induction instruments
Induction instruments require alternating fluxes. Two fluxes displaced in time and space induce eddy currents in a conducting disc, producing average torque approximately proportional to
Td\propto\Phi1\Phi2\sin\alpha
where \Phi1 and \Phi2 are flux magnitudes and \alpha is their phase displacement. Traditional induction energy meters integrate power through disc rotation. They do not operate on DC and are affected by frequency, temperature, friction and waveform distortion.
2.5 Electrostatic instruments
Electrostatic instruments use the force between charged electrodes. For capacitance C(\theta), stored energy is CV2/2, giving
Td=\frac{1}{2}V2\frac{dC}{d\theta}
They draw negligible steady current and can measure high AC or DC voltages. Their low torque, susceptibility to stray electric fields and bulky construction restrict general-purpose use.
Instrument
Operating quantity
AC/DC capability
Scale
Typical application
PMMC
Current in magnetic field
DC; AC with rectifier
Uniform
Sensitive DC indication
Moving iron
Square of current
AC and DC
Nonlinear
Industrial ammeters and voltmeters
Electrodynamic
Product of two currents
AC and DC
Application-dependent
Power measurement
Induction
Interaction of AC fluxes
AC only
Nearly uniform by design
Energy measurement
Electrostatic
Square of voltage
AC and DC
Nonlinear
High-voltage measurement
3. Instrument Errors and Loading
An ammeter inserted in series should ideally have zero resistance. Its actual burden resistance reduces circuit current and dissipates power. A voltmeter connected in parallel should ideally have infinite resistance. If a source with Thévenin resistance Rth is measured by a voltmeter of resistance Rv, then
Vindicated=Vth\frac{Rv}{Rth+Rv}
The loading error becomes significant when Rv is not much greater than Rth. Analog instruments also exhibit friction, zero shift, parallax, temperature effects, magnetic aging and calibration drift. Mirror scales reduce parallax by requiring the pointer to obscure its reflection.
Accuracy class normally specifies permissible error as a percentage of full-scale value. A class 1 instrument with a 100 V range may have an error of ±1 V anywhere on the scale. At a 10 V reading this corresponds to ±10% of the indicated value, illustrating why measurements should preferably be made in the upper part of a range.
4. General Four-Arm Bridge
A bridge compares four impedances. Adopt the following unambiguous convention:
Z1 connects A–B;
Z2 connects B–C;
Z3 connects C–D;
Z4 connects D–A;
the source is connected between A and C;
the null detector is connected between B and D.
Correct four-arm bridge with source across A and C and detector across B and D
Figure: Correct four-arm bridge with source across A and C and detector across B and D
Figure: Four-arm bridge with four distinct perimeter impedances, source A–C and null detector B–D.
With an ideal open-circuit detector, the divider voltages relative to C are
This complex equation requires both real and imaginary components to balance in an AC bridge. The relation assumes linear impedances, a source connected only across A–C and a detector drawing negligible current at null.
Away from balance, the open-circuit detector voltage is
Actual detector current also depends on the bridge's Thévenin impedance and detector impedance. Thus a mathematically balanced bridge may still be difficult to adjust if detector sensitivity is inadequate.
5. DC Resistance Bridges
5.1 Wheatstone bridge
For four resistive arms, let Z1=Rx, Z2=R2, Z3=R3 and Z4=R4. The balance equation gives
Rx=\frac{R2R4}{R3}
The Wheatstone bridge is most effective for medium resistances. At low resistance, lead and contact resistances become comparable with the unknown. At high resistance, leakage, insulation and detector-current limitations dominate.
Worked Example 1: Wheatstone bridge
A bridge balances with R2=100\ \Omega, R3=250\ \Omega and R4=500\ \Omega. Therefore,
Rx=\frac{100\times500}{250}=200\ \Omega
If the standard resistors are uncorrelated, the first-order relative standard uncertainty is
Correlations must be included when standards share calibration or environmental influences.
5.2 Kelvin double bridge
The Kelvin double bridge measures very low resistance by separating current-carrying and potential connections. Consider the conventional topology in which:
Rx, link resistance r and standard S form the main low-resistance path;
P and Q are the main ratio arms across the outer current terminals;
p and q form an auxiliary divider across the link;
the detector connects the two divider junctions.
For this stated topology, ideal detector balance gives
All of P,Q,p,q,r,S are resistances. The second term explicitly includes the auxiliary arms and link resistance; it must not be omitted in the unmatched case. When the ratios are matched,
\frac{P}{Q}=\frac{p}{q}
and the link correction cancels:
Rx=\frac{P}{Q}S
This cancellation is the central advantage of the Kelvin bridge. Four-terminal connections also prevent voltage-sensing leads from including the principal current-lead voltage drops.
6. AC Bridges
AC bridges compare complex impedances and normally require two independent adjustments. The source must be stable in frequency and waveform, while the detector must respond selectively at the measurement frequency.
6.1 Schering bridge
The Schering bridge measures capacitance and dielectric loss. Use the general bridge convention with:
Here Cx and Rx form the series model of the unknown lossy capacitor, Cs is a nominally loss-free standard capacitor, R4 is in parallel with C4, and R3 is noninductive.
Multiplying through by 1+j\omega R4C4 and equating real and imaginary components yields
Cx=Cs\frac{R4}{R3}, \qquad Rx=R3\frac{C4}{Cs}
The units are consistent: the first relation gives farads and the second ohms. For the series-loss model, the dissipation factor is
\tan\delta=\omega CxRx=\omega R4C4
These equations apply to the stated arm arrangement and series equivalent model. Different bridge labeling or a parallel loss model requires a fresh derivation rather than relabeling formulas mechanically.
Worked Example 2: Schering bridge
At 1 kHz, a bridge balances with Cs=100\ pF, R3=1.00\ k\Omega, R4=10.0\ k\Omega and C4=100\ pF. Then
Thus the dielectric dissipation factor is approximately 0.628% at 1 kHz.
6.2 Wien bridge
A Wien bridge is used for frequency measurement and oscillator stabilization. For a series arm Z1=R1+1/(j\omega C1), a parallel arm Z4=R4/(1+j\omega R4C4) and resistive arms Z2=R2 and Z3=R3, balance requires
f=\frac{1}{2\pi\sqrt{R1R4C1C4}}
and
\frac{R2}{R3}=\frac{R1}{R4}+\frac{C4}{C1}
For the symmetrical case R1=R4=R and C1=C4=C,
f=\frac{1}{2\pi RC}, \qquad \frac{R2}{R3}=2
Other important AC bridges include Maxwell and Hay bridges for inductance, Anderson bridges for precision inductance comparison, and transformer-ratio bridges for high-accuracy impedance metrology.
7. Sensitivity, Parasitics and Uncertainty
Bridge sensitivity is the detector response per unit fractional change in the unknown. High source voltage may improve sensitivity but can cause self-heating, dielectric stress or magnetic nonlinearity. Sensitivity must therefore be optimized rather than maximized without limit.
Major parasitic effects include:
lead and contact resistance in low-resistance measurements;
leakage conductance in high-resistance and capacitance measurements;
lead inductance and stray capacitance at high frequency;
mutual coupling between bridge arms;
thermoelectric EMFs in DC microvolt circuits;
harmonic content in the source;
detector phase response and finite input impedance.
Guarding holds leakage-sensitive surfaces near the potential of the measurement node so that leakage current bypasses the detector. Shielding reduces electric and magnetic interference, while twisted pairs and four-terminal connections reduce inductive pickup and lead-resistance error.
A defensible uncertainty budget should include standard calibration, ratio error, detector resolution, repeatability, temperature coefficients, residual parasitics, frequency uncertainty and model inadequacy. A null reading does not imply zero uncertainty.
8. Applied Case: Cable Insulation Acceptance Test
Background. This is an illustrative case. A utility laboratory receives a medium-voltage cable whose insulation capacitance and dielectric loss must be verified before commissioning.
Problem. A conventional handheld capacitance meter can estimate capacitance but lacks traceable dissipation-factor accuracy at the specified test frequency. Surface leakage and electromagnetic interference are also significant.
Analysis. The laboratory selects a Schering bridge because it compares the cable directly with a calibrated low-loss capacitor. The cable is represented by a series equivalent capacitance and loss resistance at the test frequency. The bridge is arranged exactly as specified in Section 6.1. Shielded leads are used, the cable screen is connected to the guard system, and the bridge enclosure is bonded to protective earth.
Approach. The test team first verifies source frequency and bridge insulation. Voltage is increased gradually within the equipment and cable ratings. The resistive and capacitive controls are alternately adjusted because they are not perfectly orthogonal. Several balance readings are taken after thermal stabilization. Capacitance and dissipation factor are calculated from
Cx=Cs\frac{R4}{R3}, \qquad \tan\delta=\omega R4C4
An uncertainty budget includes standard-capacitor calibration, resistor ratio, frequency, repeatability and residual leakage.
Outcome. The measured capacitance agrees with the cable design value, while the dissipation factor remains below the contractual limit. Repetition at increased voltage shows no abnormal rise, so the cable passes the electrical acceptance criterion.
Learning points. Correct topology is essential because a wrong arm assignment invalidates the equations. Guarding and shielding can be as important as bridge arithmetic. Dielectric loss must be reported with frequency, voltage, temperature, equivalent-circuit model and uncertainty.
9. Safety, Governance and Contemporary Practice
High-voltage bridges require interlocked enclosures, visible grounding procedures, current limiting and discharge of capacitive specimens before access. Laboratories should maintain calibration traceability, controlled procedures, equipment histories and documented uncertainty in accordance with applicable quality systems such as ISO/IEC 17025.
Modern automatic bridges use digital synthesis, synchronous detection and numerical balancing. Impedance analyzers measure both magnitude and phase over wide frequency ranges, but their results still depend on bridge concepts, fixture compensation and equivalent-circuit selection. Machine-learning methods may assist drift detection or anomaly recognition, but they should not replace traceable calibration or physically justified uncertainty analysis.
10. Measurement Best Practices
Define the measurand, frequency, waveform and equivalent-circuit model before testing.
Confirm bridge arm labels and derive the balance equation from the actual connections.
Select standards close to the unknown to maximize sensitivity.
Use four-terminal methods for low resistance and guarding for high impedance.
Minimize self-heating and allow thermal stabilization.
Calibrate at conditions representative of use.
Repeat measurements after reversing polarity where thermal EMFs are relevant.
Record environmental conditions, uncertainty and traceability.
Never infer high accuracy solely from a sharp null.
Chapter Summary
Classical analog instruments establish indication through the balance of deflecting and controlling torque, while damping governs transient response. PMMC instruments provide sensitive linear DC measurement; moving-iron instruments operate on AC and DC; electrodynamic instruments measure power; induction instruments use AC flux interaction; and electrostatic instruments are useful for high voltage.
A four-arm bridge with source A–C and detector B–D balances when Z1Z3=Z2Z4. The Wheatstone bridge measures medium resistance, while the Kelvin double bridge suppresses link and lead resistance when its main and auxiliary ratios are matched. The Schering bridge determines capacitance and dielectric loss, and the Wien bridge provides frequency-selective balance. Reliable measurements require correct topology, adequate detector sensitivity, control of parasitic effects, traceable standards and realistic uncertainty evaluation.
Key Terms
Analog instrument: Instrument providing a continuous mechanical indication of a measured quantity.
Balance condition: Relationship among bridge arms that produces zero detector voltage.
Burden: Electrical loading introduced by a measuring instrument.
Controlling torque: Restoring torque that establishes steady pointer position.
Critical damping: Damping that gives the fastest nonoscillatory response.
Deflecting torque: Input-dependent torque that moves an indicating element.
Dielectric loss: Energy dissipated in an insulating material under an alternating field.
Dissipation factor: Ratio characterizing dielectric loss, commonly expressed as \tan\delta.
Electrodynamic instrument: Instrument producing torque through interaction of fixed- and moving-coil fields.
Four-terminal measurement: Method separating current and voltage connections to reduce lead-resistance error.
Guarding: Technique that diverts leakage currents away from sensitive measurement nodes.
Kelvin double bridge: Bridge for precision measurement of very low resistance.
Loading error: Change in the measured circuit caused by the measuring instrument.
Moving-iron instrument: Instrument using magnetic attraction or repulsion of soft iron.
Null detector: Sensitive device used to identify bridge balance.
PMMC instrument: Permanent-magnet moving-coil instrument primarily used for DC measurement.
Schering bridge: AC bridge for measuring capacitance and dielectric loss.
Wheatstone bridge: Four-resistor bridge for resistance comparison.
Wien bridge: Frequency-sensitive RC bridge.
Review Questions
Short Answer Questions
Distinguish between controlling torque and damping torque.
Why does an unrectified PMMC instrument not indicate sinusoidal AC?
State the general balance condition for the bridge convention used in this chapter.
Why are four-terminal connections required for low-resistance measurements?
What physical quantity is represented by the dissipation factor of a capacitor?
Long Answer Questions
Derive the dynamic equation of an analog indicating instrument and discuss the effects of underdamping, critical damping and overdamping.
Compare PMMC, moving-iron, electrodynamic, induction and electrostatic instruments with respect to torque, scale, applications and errors.
Derive the Wheatstone bridge balance equation and discuss sensitivity, loading and uncertainty.
Explain the Kelvin double bridge, including the role of its auxiliary ratio arms and link-resistance correction.
Derive the Schering bridge balance equations for the arm arrangement specified in this chapter.
Analytical / Application Questions
Design a PMMC voltmeter range extension for a 1 mA, 100 Ω movement to measure 100 V, and calculate its input resistance.
A Kelvin bridge has unequal main and auxiliary ratios. Use the full balance equation to quantify the link-resistance correction for selected component values.
Develop an uncertainty budget for a Schering-bridge measurement of a high-voltage cable, including calibration, frequency, leakage and repeatability contributions.
Multiple Choice Questions with Answers
Question 1
For the bridge convention in which Z1 connects A–B, Z2 connects B–C, Z3 connects C–D, Z4 connects D–A, the source is A–C and the detector is B–D, what is the balance condition?
A. Z1Z2=Z3Z4
B. Z1Z3=Z2Z4
C. Z1Z4=Z2Z3
D. Z1+Z3=Z2+Z4
Correct Answer: B. Z1Z3=Z2Z4
Question 2
For the stated bridge topology, the open-circuit detector voltage VBD is which expression?
A. Vs\left(\frac{Z2}{Z1+Z2}-\frac{Z3}{Z4+Z3}\right)
B. Vs\left(\frac{Z1}{Z1+Z2}-\frac{Z3}{Z4+Z3}\right)
C. Vs\frac{Z1+Z3}{Z2+Z4}
D. Vs\frac{Z1Z3}{Z2Z4}
Correct Answer: A. Vs\left(\frac{Z2}{Z1+Z2}-\frac{Z3}{Z4+Z3}\right)
Question 3
Why does an ideal unrectified PMMC instrument produce no steady indication for a symmetrical sinusoidal current?
A. Its control spring blocks alternating current
B. Its coil inductance is infinite
C. Its torque reverses with current and has zero average
D. Its torque is proportional to current squared
Correct Answer: C. Its torque reverses with current and has zero average
Question 4
The deflecting torque of a moving-iron instrument is proportional to which quantity under its energy model?
A. i only
B. i2(dL/d\theta)
C. di/dt only
D. 1/i
Correct Answer: B. i2(dL/d\theta)
Question 5
A Wheatstone bridge has R2=100\ \Omega, R3=250\ \Omega and R4=500\ \Omega. What unknown resistance balances the bridge?
A. 50 Ω
B. 125 Ω
C. 200 Ω
D. 1.25 kΩ
Correct Answer: C. 200 Ω
Question 6
What occurs in a Kelvin double bridge when P/Q=p/q?
A. The standard resistance is short-circuited
B. The link-resistance correction term cancels
C. The detector must carry the main current
D. The bridge becomes frequency-selective
Correct Answer: B. The link-resistance correction term cancels
Question 7
A Schering bridge has Cs=100\ pF, R3=1\ k\Omega and R4=10\ k\Omega. For the topology stated in the chapter, what is Cx at balance?
A. 10 pF
B. 100 pF
C. 1 nF
D. 10 nF
Correct Answer: C. 1 nF
Question 8
In a symmetrical Wien bridge with R1=R4=R and C1=C4=C, the balance frequency is:
A. 1/(RC)
B. 1/(2\pi RC)
C. RC/(2\pi)
D. 2\pi RC
Correct Answer: B. 1/(2\pi RC)
Question 9
A voltmeter of resistance Rv measures a source with Thévenin voltage Vth and resistance Rth. What voltage is indicated?
A. Vth(Rth+Rv)/Rv
B. VthRth/(Rth+Rv)
C. VthRv/(Rth+Rv)
D. Vth(Rv-Rth)
Correct Answer: C. VthRv/(Rth+Rv)
Question 10
In an electrodynamic wattmeter supplied with sinusoidal voltage and current, average torque is proportional to:
A. Reactive power only
B. VI\cos\phi
C. V/I
D. I2/V
Correct Answer: B. VI\cos\phi